    Design and testing of an MRI compatible therapeutic transducer

  • Author

    Brosch, Jared ; Morris, G. Adam ; Wilson, Tommy ; Talavage, Thomas

  • Author_Institution
    Etalon a PiezoTechnol. Co., Lebanon, IN, USA
  • Volume
    2
  • fYear
    2001
  • fDate
    2001
  • Firstpage
    1169
  • Abstract
    Testing of many different piezoelectric elements in an MRI has shown that all are safe except for the Keramos NOVA-7A, which is not acceptable for imaging. The greatest improvement in the design of an MRI compatible transducer is the use of a "delay line" to remove the transducer out of the immediate field of view. Finally, functional MRI testing shows that the Exogen therapeutic transducer may work by increasing blood flow in the region of its use
  • Keywords
    biomedical MRI; biomedical transducers; biomedical ultrasonics; delay lines; radiation therapy; Exogen transducer; Keramos NOVA-7A; MRI compatible therapeutic transducer; blood flow; delay line; functional MRI testing; medical transducers; piezoelectric elements; ultrasonic transducers; Biochemistry; Biomedical imaging; Biomedical transducers; Bones; Imaging phantoms; Magnetic resonance imaging; Piezoelectric transducers; Plastics; Testing; Ultrasonic transducers;