Title :
A digital envelope detection filter for blood pressure measurement
Author :
Lee, J.Y. ; Kim, J.K. ; Yoon, G.
Author_Institution :
Med. Application Team, Samsung Adv. Inst. of Technol., Yongin, South Korea
Abstract :
We developed a finger type blood pressure measurement system using a volume oscillometric algorithm. The digital envelope detection filter (EDF) method was applied to volume oscillometric signals which were acquired from an air-filled occlusive finger cuff. In the case of using the EDF, we could reduce the difference error and standard deviation by 30-40%.
Keywords :
blood pressure measurement; blood vessels; digital filters; medical signal detection; patient monitoring; air-filled occlusive finger cuff; blood pressure measurement; blood pressure monitoring; difference error; digital envelope detection filter; finger artery; standard deviation; volume oscillometric signals; Arteries; Biomedical monitoring; Blood pressure; Digital filters; Envelope detectors; Fingers; Finite impulse response filter; Frequency; Pressure measurement; Volume measurement;
Conference_Titel :
Engineering in Medicine and Biology Society, 2001. Proceedings of the 23rd Annual International Conference of the IEEE
Print_ISBN :
0-7803-7211-5
DOI :
10.1109/IEMBS.2001.1018896