DocumentCode :
3850408
Title :
Alignment Procedure Used in the BIPM Watt Balance
Author :
Estefania de Mirand?s;Hao Fang;Adrien Kiss;Stéphane Solve;Michael Stock;Alain Picard
Author_Institution :
Bureau International des Poids et Mesures, Sè
Volume :
60
Issue :
7
fYear :
2011
Firstpage :
2415
Lastpage :
2421
Abstract :
The Bureau International des Poids et Mesures (BIPM) is developing a watt balance to measure accurately the Planck constant h by linking it to the kilogram. The kilogram is currently the only base unit of the International System of Units still defined by a material artifact. A fixed numerical value of h could be the basis for the expected future redefinition of the kilogram in terms of fundamental constants. One main concern of the watt balance experiment is the alignment of the involved vectors. In particular, the electromagnetic force exerted on the watt balance coil needs to be carefully aligned parallel to the gravitational force exerted on the test mass. To address this requirement, we have developed an original experimental procedure that has been validated with a test coil which is a prototype for the ultimate watt balance coil. The result is accurate enough to enable a determination of the Planck constant with a relative uncertainty of about 1·10-8 , which is the ultimate goal of the BIPM watt balance.
Keywords :
"Coils","Solenoids","Voltage measurement","Magnetic recording","Windings","Current measurement","Lead"