• DocumentCode
    393619
  • Title

    Study of the immersion characteristic of thermocouples with a heat-pipe up to 960°C

  • Author

    Gotoh, Masahiko

  • Author_Institution
    Tamagawa Univ., Tokyo, Japan
  • Volume
    1
  • fYear
    2002
  • fDate
    5-7 Aug. 2002
  • Firstpage
    503
  • Abstract
    In this paper results of the measurements on the irregular thermal EMF by means of the thermoelectric scanning are reported. Scanning is carried out in the pressure controlled heat pipe at 960°C over 50 cm along the thermocouples. Measurement was performed with an Au/Pt, two Pt/Pd and a multi-element wire thermocouples. One of the Pt/Pd thermocouples was subjected to 500 hours exposure to 1300°C air environment. The other Pt/Pd thermocouple is a newly fabricated one and act as the reference. The study found no significant difference in the spurious Emf between the two Pt/Pd thermocouples. Discrepancy due to the different immersion was 55 mK for the two Pt/Pd at most. This result suggests that they are superior to the type S and R thermocouples in terms of the immersion characteristics. It is very important to access the source of uncertainty of thermocouples due to the different immersion profile of the furnace. The study provides more realistic information for this purpose.
  • Keywords
    gold; heat pipes; palladium; platinum; temperature distribution; temperature measurement; thermocouples; 1300 degC; 500 hr; Au; Pd; Pt; furnace; heat pipe; immersion profile; insertion error; irregular thermal EMF; multiple element wire thermocouples; thermoelectric scanning; Gold; Performance evaluation; Petroleum; Platinum; Pressure control; Temperature control; Temperature distribution; Temperature measurement; Thermoelectricity; Wire;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    SICE 2002. Proceedings of the 41st SICE Annual Conference
  • Print_ISBN
    0-7803-7631-5
  • Type

    conf

  • DOI
    10.1109/SICE.2002.1195455
  • Filename
    1195455