DocumentCode
393623
Title
Estimation of pressure head and flow rate in a continuous-flow artificial heart-in vivo evaluation
Author
Tanaka, A. ; Yoshizawa, M. ; Aizawa, Y. ; Olegario, P. ; Abe, K. ; Yambe, T. ; Nitta, S.
Author_Institution
Graduate Sch. of Eng., Tohoku Univ., Sendai, Japan
Volume
1
fYear
2002
fDate
5-7 Aug. 2002
Firstpage
527
Abstract
Instead of direct measurement of pressure head and flow rate of a centrifugal blood pump, the authors previously proposed a method of estimating these measurements. In this study, to ascertain the adequacy of the proposed method, an acute animal experiment was carried out using an adult goat equipped with an artificial heart consisting of two centrifugal pumps. The results indicated that the ARX model identified even in a mock circulation could estimate flow rate in the recipient accurately enough, although estimation accuracy of pressure head was not high.
Keywords
artificial organs; blood pressure measurement; flow measurement; ARX model; acute animal experiment; centrifugal blood pump; continuous-flow artificial heart; flow rate estimation; goat; pressure head estimation; total artificial heart; Aging; Animals; Artificial heart; Blood; Cancer; Fluid flow measurement; Power supplies; Pressure measurement; Viscosity; Voltage;
fLanguage
English