    Evaluation of a pressure-controlled water heat pipe for accurate comparison of platinum resistance thermometers

  • Author

    Tamba, J. ; Kishimoto, I. ; Arai, M.

  • Author_Institution
    Nat. Inst. of Adv. Ind. Sci. & Technol., Nat. Metrol. Inst. of Japan, Tsukuba, Japan
  • Volume
    4
  • fYear
    2002
  • fDate
    5-7 Aug. 2002
  • Firstpage
    2435
  • Abstract
    A pressure-controlled water heat pipe was evaluated as a comparison furnace for Standard Platinum Resistance Thermometers (SPRTs). Its calibration uncertainties in the range from 65°C to 157°C were estimated. In direct comparison of SPRTs with a calibrated SPRT, the combined uncertainty of the comparison is estimated to be 0.081 mK. On using the heat pipe as a calibration facility, the combined uncertainty of the calibration is estimated to be 3.1 mK.
  • Keywords
    calibration; heat pipes; measurement uncertainty; platinum; pressure control; resistance thermometers; Pt; calibration facility; calibration uncertainties; comparison furnace; pressure-controlled water heat pipe; standard Pt resistance thermometers; thermometry; Calibration; Electrical resistance measurement; Furnaces; Platinum; Pressure control; Resistance heating; Temperature; Thermal resistance; Uncertainty; Water heating;