Abstract
The applicability of the time-reversal theory by the pseudospectral time-domain method is investigated to localize a medical implant equipped with an ultrawideband radio and embedded in a randomly lossy dielectric medium. A localization algorithm is implemented and tested for various test cases. A comparative analysis with the finite difference time-domain method implementation is performed in terms of memory occupation and computational time.
Keywords
biological tissues; finite difference time-domain analysis; prosthetics; ultra wideband communication; UWB source localization; biological tissues; finite difference time-domain method; medical implant; memory occupation; pseudospectral time-domain time-reversal method; random lossy dielectric medium; ultrawideband radio; Conductivity; Electromagnetic scattering; Finite difference methods; Media; Observers; Permittivity; Time-domain analysis; Finite difference time-domain (FDTD); locating system; pseudospectral method; time reversal (TR); ultrawideband (UWB) technology;