• DocumentCode
    416465
  • Title

    The simple monitoring method of benzene using quartz crystal microbalance type sensor

  • Author

    Noda, K. ; Naganawa, R. ; Tao, H.

  • Author_Institution
    Nat. Inst. of Adv. Ind. Sci. & Technol., Onogawa, Japan
  • Volume
    1
  • fYear
    2003
  • fDate
    4-6 Aug. 2003
  • Firstpage
    62
  • Abstract
    Highly sensitive and selective monitoring method of benzene was newly developed. Selective reaction reagent that was used in the gas detecting tube was utilized. Using diiodine pentaoxide (I/sub 2/O/sub 5/), iodine (I/sub 2/) in proportion to the benzene concentration was formed. This was adsorbed on the silver electrode, and increase of the weight was measured. In case of 100 ppb of benzene concentration and fundamental frequencies of 9 MHz at 30/spl deg/C, the frequency response of about -370 Hz was obtained. As a result of the examination, it was found that the minimum detection resolution was 0.3 ppb/Hz.
  • Keywords
    air pollution measurement; chemical reactions; computerised monitoring; gas sensors; iodine compounds; organic compounds; quartz; -370 Hz; 30 degC; 9 MHz; benzene monitoring; diiodine pentaoxide; gas detecting tube; quartz crystal microbalance; reaction reagent;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    SICE 2003 Annual Conference
  • Conference_Location
    Fukui, Japan
  • Print_ISBN
    0-7803-8352-4
  • Type

    conf

  • Filename
    1323315