    In vitro test of an adaptive flow controller for a continuous flow LVAD

  • Author

    Wu, Yi ; Allaire, Paul ; Tao, Gang ; Liu, Yingjie

  • Author_Institution
    Dept. of Mech. Eng., Virginia Univ., Charlottesville, VA, USA
  • Volume
    2
  • fYear
    2004
  • fDate
    June 30 2004-July 2 2004
  • Firstpage
    1647
  • Abstract
    An adaptive flow controller designed for a magnetically levitated continuous flow (CF) left ventricular assist device (LVAD) was implemented on a real-time prototype controller board and put in a mock human circulatory loop for an in vitro test. Test results showed that the designed controller maintained the key haemodynamic variable of a congestive heart failure (CHF) patient within the normal physiologic range in either specified rest or exercise conditions. The transient response is slow and need improvement.
  • Keywords
    adaptive control; biomedical equipment; cardiology; control system synthesis; flow control; haemodynamics; magnetic levitation; transient response; adaptive flow controller design; congestive heart failure; continuous flow left ventricular assist device; haemodynamic variable; in vitro test; magnetic levitation; mock human circulatory loop; real time prototype controller board; transient response;