Abstract :
A hardware implementation is presented for receive beamforming of raw data collected from a commercial ultrasound scanner. The hardware is built with programmable components, FPGA and microcontroller, implemented in a standalone board configurable from a PC and connected to the ultrasound scanner. An application to receive data beamforming for the transverse component estimation in elasticity imaging is presented. Results show that the mean standard deviation of the transverse estimate is equal to 0.03 times the lateral wavelength. It can be compared to a mean standard deviation of the axial estimate of 0.02 time the axial wavelength. Results emphasize the interest of receive data beamforming for providing lateral RF signals applied to elasticity imaging.
Keywords :
array signal processing; biomedical ultrasonics; elasticity; field programmable gate arrays; microcontrollers; ultrasonic arrays; ultrasonic imaging; FPGA; commercial ultrasound scanner; lateral RF signals; microcontroller; programmable components; receive beamforming; transverse elasticity imaging; Array signal processing; Elasticity; Hardware; High-resolution imaging; Low pass filters; Parameter estimation; Probes; RF signals; Radio frequency; Ultrasonic imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium, 2004 IEEE
ISSN :
1051-0117
Print_ISBN :
0-7803-8412-1
Type :
conf
DOI :
10.1109/ULTSYM.2004.1418084
Filename :
1418084
Link To Document :
بازگشت