• DocumentCode
    438504
  • Title

    Noise properties of four strategies for incorporation of scatter and attenuation information in PET reconstruction

  • Author

    Tamal, M. ; Reader, A.J. ; Markiewicz, P.J. ; Hastings, D.L. ; Julyan, P.J.

  • Author_Institution
    Sch. of Chem. Eng. & Anal. Sci., Manchester Univ.
  • Volume
    5
  • fYear
    2004
  • fDate
    16-22 Oct. 2004
  • Firstpage
    2840
  • Lastpage
    2844
  • Abstract
    Conventional methods for dealing with attenuation and scatter can degrade the reconstructed image quality, particularly if the attenuating medium is large (as in whole body 3D PET). In such cases, a substantial scatter subtraction is performed followed by amplification of the remaining data (to correct for attenuation), which results in noisy reconstructions. More recent methods used with iterative reconstruction include the attenuation in the system model in conjunction with either pre-scatter subtraction or a separate addition of the scatter component after each application of the forward model. This work compares these more conventional approaches to including attenuation and scatter within the EM algorithm with a fully unified scatter and attenuation model - whereby all attenuation and scattering effects are included within the system matrix. For this case all acquired data are used and regarded as information by the reconstruction algorithm. This work indicates that for a large attenuating medium there are notable differences between the four ways of including attenuation and scatter within the reconstruction - fully pre-correction of the data is inferior compared to all the other methods. For the case of simple shift-invariant Gaussian model of scatter - subtraction, additive and unified model methods shows similar variance-bias characteristics. Multiple realisations of simulated data sets have been used to compare the performance of this model with other methods.
  • Keywords
    image reconstruction; iterative methods; medical image processing; noise; positron emission tomography; 3D PET; EM algorithm; Gaussian model; PET reconstruction; additive approach; attenuation information; fully precorrected method; image quality; image reconstruction; iterative reconstruction; noise properties; scatter component; subtraction method; unified model; variance-bias characteristics; Attenuation; Degradation; Image quality; Image reconstruction; Iterative algorithms; Iterative methods; Positron emission tomography; Reconstruction algorithms; Scattering; Whole-body PET;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2004 IEEE
  • Conference_Location
    Rome
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-8700-7
  • Electronic_ISBN
    1082-3654
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2004.1466279
  • Filename
    1466279