• DocumentCode
    438609
  • Title

    Misalignment between emission scan and X-ray CT derived attenuation maps causes artificial defects in attenuation corrected myocardial perfusion SPECT

  • Author

    Fricke, Harald ; Fricke, Eva ; Weise, Reiner ; Kammeier, Annett ; Lindner, Oliver ; Burchert, Wolfgang

  • Author_Institution
    Inst. of Molecular Biophys., Radiopharmacy & Nuclear Med., Heart & Diabetes Center North Rhine-Westphalia, Bad Oeynhausen, Germany
  • Volume
    6
  • fYear
    2004
  • fDate
    16-22 Oct. 2004
  • Firstpage
    3422
  • Abstract
    Attenuation correction increases the specificity of myocardial perfusion SPECT. However, we noticed an unusually high number of scans with defects only visible in the attenuation-corrected images. We suspected these to be false positive readings. Visual inspection using the supplied software suggested mismatch in the ventrodorsal (Y-) direction between SPECT images and transmission maps as an explanation. As the fusion tool only allows for a coarse grading, we wrote software to quantify the mismatch. A phantom study was done to verify that the observed mismatch can cause the defect patterns visible in the attenuation-corrected images. 25 patients who showed the most pronounced artifact were chosen for re-alignment and re-evaluation. Overall, the defects in the attenuation-corrected images got less intense (15/25) or even vanished (6/25) after re-aligning emissions images and transmission maps. In response to our complaints, the vendor replaced the support rollers which should prevent the bed from deflecting with a re-engineered, more robust version. No more clinically relevant artefacts were observed after this modification. Evaluation of another 28 probably-normal patients showed the mean mismatch between emission and transmission scan to be significantly reduced. We conclude that great care has to be taken to ensure correct alignment of the scans even in a dual-modality imaging device. Bed deflection can be a major source for misalignment and artifacts.
  • Keywords
    cardiovascular system; diagnostic radiography; haemorheology; medical computing; muscle; phantoms; single photon emission computed tomography; X-ray CT; artificial defects; attenuation maps; attenuation-corrected image; bed deflection; dual-modality imaging device; emission scan; myocardial perfusion SPECT; phantom study; visual inspection; Attenuation; Computed tomography; Image analysis; Imaging phantoms; Inspection; Myocardium; Robustness; Single photon emission computed tomography; Stress; X-ray imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2004 IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-8700-7
  • Electronic_ISBN
    1082-3654
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2004.1466622
  • Filename
    1466622