• DocumentCode
    438669
  • Title

    Exact image reconstruction for selected ROI in helical cone-beam CT

  • Author

    Pan, Xiaochuan ; Zou, Yu ; Xia, Dan ; Sidky, Emil ; Jiang, Tianzi

  • Author_Institution
    Dept. of Radiol., Chicago Univ., IL, USA
  • Volume
    6
  • fYear
    2004
  • fDate
    16-22 Oct. 2004
  • Firstpage
    3773
  • Abstract
    The sufficiency condition(s) are derived for exact image reconstruction of a 3D ROI from projections acquired with a reduced helical scan over an angular range considerably smaller than that required by image reconstruction in, e.g., the conventional long object problem, in which the scanned angular range is often more than 2π. ROI reconstruction is investigated by a recently developed filtered-backprojection algorithm that can make use of data acquired with a reduced helical scan. Preliminary numerical studies demonstrate and validate the ROI reconstruction. This work may have significant practical implications because a reduced scan in CT often translates to reduced motion artifacts and reduced radiation dose delivered to the subject.
  • Keywords
    computerised tomography; data acquisition; dosimetry; image reconstruction; image scanners; medical image processing; computed tomography; data acquisition; filtered-backprojection algorithm; helical cone-beam CT; helical scan; image reconstruction; radiation dose; region-of-interest; Automation; Computed tomography; Detectors; Image reconstruction; Image segmentation; Joining processes; Radiology; Sufficient conditions;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2004 IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-8700-7
  • Electronic_ISBN
    1082-3654
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2004.1466701
  • Filename
    1466701