• DocumentCode
    462337
  • Title

    Position Determination and Resolution of Position Sensitive Neutron Detectors Limited by Charge Equalization and Noise

  • Author

    Bönisch, Sven P. ; Namaschk, Bernhard ; Wulf, Friedrich

  • Author_Institution
    Hahn-Meitner-Inst. of Berlin
  • Volume
    1
  • fYear
    2006
  • fDate
    Oct. 29 2006-Nov. 1 2006
  • Firstpage
    321
  • Lastpage
    324
  • Abstract
    The conventional design of detector electronics for resistive-wire position sensitive neutron detectors (PSD) using charge sensitive preamplifiers and gauss-pulse shaping devices causes uncertainty in spatial position and resolution due to charge equalization between both ends of the resistive PSD. Errors introduced by charge equalization may be in excess of limitations given by noise, linearity, dynamic range or EMI. This paper describes in detail charge equalization effects and their drawbacks for a PSD system design. Error estimations for a conventional detector design are given. Attained results should help to better understand PSD electronics design and to provide data for system optimization.
  • Keywords
    electromagnetic interference; neutron detection; nuclear electronics; position sensitive particle detectors; Gauss pulse shaping devices; charge equalization; charge sensitive preamplifier; detector electronics design; noise; position sensitive neutron detectors; Dynamic range; Electromagnetic interference; Gaussian processes; Linearity; Neutrons; Noise shaping; Position sensitive particle detectors; Preamplifiers; Spatial resolution; Uncertainty;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2006. IEEE
  • Conference_Location
    San Diego, CA
  • ISSN
    1095-7863
  • Print_ISBN
    1-4244-0560-2
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2006.356166
  • Filename
    4179005