DocumentCode
462605
Title
System Integration of the LabPET Small Animal PET Scanner
Author
Tetrault, M.-A. ; Viscogliosi, N. ; Riendeau, J. ; Belanger, F. ; Michaud, J.-B. ; Semmaoui, H. ; Berard, P. ; Lemieux, F. ; Arpin, L. ; Cadorette, J. ; Pepin, C.M. ; Robert, G. ; Lepage, M.D. ; Lecomte, R. ; Fontaine, R.
Author_Institution
Dept. of Electr. & Comput. Eng., Univ. de Sherbrooke, Que.
Volume
3
fYear
2006
fDate
Oct. 29 2006-Nov. 1 2006
Firstpage
1880
Lastpage
1884
Abstract
To address modern molecular imaging requirements, a digital positron emission tomography scanner for small animals has been developed at Universite de Sherbrooke. Based on individual readout of avalanche photodiodes (APD) coupled to a LYSO/LGSO phoswich array, the scanner supports up to 3072 channels in a 16.2 cm diameter, 7.5 cm axial field of view with an isotropic 1.2 mm FWHM intrinsic spatial resolution at the center of the FOV. Custom data acquisition boards sample APD signals at 45 MHz and compute in real time crystal identification, energy and timing information of detected events at rates of up to 1250 raw counts per second per mm2 (10k cps/channel). Real time digital signal analysis also filters out events outside the photopeak with crystal granularity to eliminate Compton events and electronic noise. Retained events are then merged into a single stream through a real-time sorting tree, at which end the prompt and delayed coincidences are extracted. A single Firewire link handles both control and data transfers with a computer. The LabPETtrade features four data recording modes, giving the user the choice to retain data for research or to minimize file size for high coincidence count rate and imaging purposes. The electronic system also supports time synchronized data insertion for flags such as vital signs used in gated image reconstruction. Aside from data acquisition, hardware can generate live energy and discrimination histograms suitable for fast, automatic channel calibration.
Keywords
biomedical equipment; medical computing; positron emission tomography; 16.2 cm; 45 MHz; Firewire link; LabPET; automatic channel calibration; avalanche photodiodes; data acquisition boards; data recording; data transfers; digital positron emission tomography scanner; digital signal analysis; gated image reconstruction; phoswich array; small animal PET scanner; sorting tree; system integration; time synchronized data insertion; Animals; Avalanche photodiodes; Data acquisition; Event detection; Firewire; Molecular imaging; Positron emission tomography; Signal processing; Spatial resolution; Timing;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2006. IEEE
Conference_Location
San Diego, CA
ISSN
1095-7863
Print_ISBN
1-4244-0560-2
Electronic_ISBN
1095-7863
Type
conf
DOI
10.1109/NSSMIC.2006.354260
Filename
4179373
Link To Document