    Tailoring PET time coincidence window using CT morphological information

  • Author

    Conti, Maurizio

  • Author_Institution
    Molecular Imaging, Siemens Med. Solution, Knoxville, TN
  • Volume
    4
  • fYear
    2006
  • fDate
    Oct. 29 2006-Nov. 1 2006
  • Firstpage
    2349
  • Lastpage
    2353
  • Abstract
    In this paper we propose a general method to reduce electronic time coincidence window in PET, based on a-priori morphological information obtained via a CT scan. This allows rejecting randoms without losing true coincidence events, improving image quality and true data throughput. Experimental results show a significant random fraction decrease and an image variance reduction.
  • Keywords
    computerised tomography; medical image processing; positron emission tomography; CT morphological information; CT scan; PET time coincidence window; electronic time coincidence window; image quality; image variance reduction; random fraction; true coincidence event; true data throughput; Attenuation; Computed tomography; Degradation; Event detection; Image quality; Nuclear and plasma sciences; Nuclear electronics; Positron emission tomography; Radioactive decay; Throughput;