• DocumentCode
    469719
  • Title

    Field of view alignment on a multimodal PET/CT scanner for small animals

  • Author

    Vaquero, JJ ; Pascau, J. ; Abella, M. ; Sisniega, A. ; Lage, E. ; Desco, M.

  • Author_Institution
    Hosp. Gen. Univ. Gregorio, Madrid
  • Volume
    4
  • fYear
    2007
  • fDate
    Oct. 26 2007-Nov. 3 2007
  • Firstpage
    2868
  • Lastpage
    2869
  • Abstract
    This work describes a procedure for the alignment of the fields of view of a multimodal system. Combined PET/CT systems for small-animal imaging usually have two axially displaced fields of view (FOV) that are mechanically aligned only to a certain degree; the remaining errors can impair sub- millimeter registrations. The correction of these residual uncertainties in the geometry is calculated using a software tool that automatically computes the misalignment between FOVs and compensates for them during image reconstruction. The procedure only requires the use of a simple phantom for calibration.
  • Keywords
    calibration; computerised tomography; image reconstruction; image registration; medical image processing; phantoms; positron emission tomography; CT; PET; calibration; field-of-view alignment; image reconstruction; multimodal scanner; phantom; residual uncertainty correction; small animal imaging; submillimeter registration; Animals; Calibration; Computed tomography; Image reconstruction; Imaging phantoms; Optical imaging; Positron emission tomography; X-ray detection; X-ray detectors; X-ray imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2007. NSS '07. IEEE
  • Conference_Location
    Honolulu, HI
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-0922-8
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2007.4436733
  • Filename
    4436733