Title :
Fully automated computer measurement of QT interval from the 12-lead electrocardiogram
Author :
Langley, P. ; Smith, F.E. ; King, S.T. ; Zheng, D. ; Haigh, A.J. ; Murray, A.
Author_Institution :
Newcastle Univ., Newcastle upon Tyne
Abstract :
The aim was to develop a fully automatic QT interval measurement algorithm for the 2006 PhysioNet/ Computers in Cardiology Challenge. The algorithm determined the Q onset and T offset points from the average beat in each lead. The QT interval measurement was calculated from the median Q onset and T offset points. Manual measurements were also made. The mean (sd) difference between automatic and reference measurements was -49 (38) ms and between manual and reference measurements was -6 (48) ms.
Keywords :
biomedical measurement; electrocardiography; medical computing; 12-lead electrocardiogram; 2006 PhysioNet/ Computers in Cardiology Challenge; ECG; QT interval; automated computer measurement; Band pass filters; Cardiology; Electrocardiography; Heart rate interval; Hospitals; Iron; Phase detection; Phase measurement; Polynomials; Q measurement;