Study on a temperature scale from 24,5561 K to 273,16 K by comparison calibration
Author
Kodaira, Kazuaki
Author_Institution
Japan Electr. Meters Inspection Corp., Japan
fYear
2009
fDate
18-21 Aug. 2009
Firstpage
3287
Lastpage
3292
Abstract
A comparison calibration apparatus using a GM cryocooler was developed. This apparatus has the structure that temperature changes of the cryocooler do not conduct directly to improve the temperature stability. A practical temperature scale from 24,5561 K to 273,16 K was examined by using the comparison calibration apparatus. Generally, a standard platinum resistance thermometer (SPRT) is calibrated at the triple points of hydrogen, neon, oxygen, argon, mercury and water in the international temperature scale of 1990 (ITS-90) for this temperature range. In this work, a capsule type SPRT was calibrated by comparison calibration method at the temperature close to these fixed point temperature. And the temperature scale using the results of the comparison calibration was compared with the temperature scale defined in the ITS-90. As a result, two temperature scales corresponded within 1,5 mK.
Keywords
calibration; cryogenics; resistance thermometers; temperature measurement; temperature scales; GM cryocooler; ITS-90; International Temperature Scale; comparison calibration apparatus; fixed point temperature; standard platinum resistance thermometer; temperature stability; Argon; Calibration; Hydrogen; Measurement standards; Platinum; Stability; Temperature distribution; Temperature measurement; Thermal resistance; Watthour meters; ITS-90; calibration; standard platinum resistance thermometer; temperature measurement;
fLanguage
English
Publisher
ieee
Conference_Titel
ICCAS-SICE, 2009
Conference_Location
Fukuoka
Print_ISBN
978-4-907764-34-0
Electronic_ISBN
978-4-907764-33-3
Type
conf
Filename
5334341
Link To Document