• DocumentCode
    504666
  • Title

    Study on a temperature scale from 24,5561 K to 273,16 K by comparison calibration

  • Author

    Kodaira, Kazuaki

  • Author_Institution
    Japan Electr. Meters Inspection Corp., Japan
  • fYear
    2009
  • fDate
    18-21 Aug. 2009
  • Firstpage
    3287
  • Lastpage
    3292
  • Abstract
    A comparison calibration apparatus using a GM cryocooler was developed. This apparatus has the structure that temperature changes of the cryocooler do not conduct directly to improve the temperature stability. A practical temperature scale from 24,5561 K to 273,16 K was examined by using the comparison calibration apparatus. Generally, a standard platinum resistance thermometer (SPRT) is calibrated at the triple points of hydrogen, neon, oxygen, argon, mercury and water in the international temperature scale of 1990 (ITS-90) for this temperature range. In this work, a capsule type SPRT was calibrated by comparison calibration method at the temperature close to these fixed point temperature. And the temperature scale using the results of the comparison calibration was compared with the temperature scale defined in the ITS-90. As a result, two temperature scales corresponded within 1,5 mK.
  • Keywords
    calibration; cryogenics; resistance thermometers; temperature measurement; temperature scales; GM cryocooler; ITS-90; International Temperature Scale; comparison calibration apparatus; fixed point temperature; standard platinum resistance thermometer; temperature stability; Argon; Calibration; Hydrogen; Measurement standards; Platinum; Stability; Temperature distribution; Temperature measurement; Thermal resistance; Watthour meters; ITS-90; calibration; standard platinum resistance thermometer; temperature measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    ICCAS-SICE, 2009
  • Conference_Location
    Fukuoka
  • Print_ISBN
    978-4-907764-34-0
  • Electronic_ISBN
    978-4-907764-33-3
  • Type

    conf

  • Filename
    5334341