• DocumentCode
    534560
  • Title

    Low-cost high-resolution animal SPECT imaging on a clinical SPECT scanner

  • Author

    Dai, Tiantian ; Ma, Tianyu ; Liu, Hui ; Cui, Junjian ; Liu, Yaqiang ; Wei, Qingyang ; Wu, Jing ; Wang, Shi ; Jin, Yongjie

  • Author_Institution
    Dept. of Eng. Phys., Tsinghua Univ., Beijing, China
  • Volume
    1
  • fYear
    2010
  • fDate
    16-18 Oct. 2010
  • Firstpage
    341
  • Lastpage
    344
  • Abstract
    In order to develop animal SPECT imaging technology with low cost, we propose a dedicated multi-pinhole collimator design scheme on a clinical SPECT scanner to achieve high resolution. Based on the proposed design one can also optimize current system design for different application purposes. This work is focused on achieving half-mm SPECT imaging resolution on a single-head clinical SPECT scanner. A seven-pinhole collimator and a four-degree-of-freedom motion control stage are assembled. High accuracy geometrical calibration is performed by scanning a point source in a hybrid orbit, which consists of three circular orbits in different axial positions and a helical orbit. The system matrix is derived from Monte-Carlo simulation and denoised by fitting each point spread function (PSF) to a 2D Gaussian function. Point source studies show a promising 0.5 mm reconstruction resolution is achievable in the studied system. Preliminary ultra micro hot rod phantom experiments demonstrate the smallest 0.75 mm hot rods are identifiable. The proposed technology is attractable for its low cost, high resolution and fine feasibility.
  • Keywords
    Monte Carlo methods; collimators; medical image processing; phantoms; single photon emission computed tomography; 2D Gaussian function; Monte-Carlo simulation; clinical SPECT scanner; four-degree-of-freedom motion control stage; half-mm SPECT imaging resolution; high accuracy geometrical calibration; low-cost high-resolution animal SPECT imaging; multi-pinhole collimator design scheme; point spread function; seven-pinhole collimator; ultramicro hot rod phantom experiments; Animals; Calibration; Collimators; Fitting; Image reconstruction; Imaging; Orbits; animal SPECT; calibration; image reconstruction; multi-pinhole collimator; system matrix;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Engineering and Informatics (BMEI), 2010 3rd International Conference on
  • Conference_Location
    Yantai
  • Print_ISBN
    978-1-4244-6495-1
  • Type

    conf

  • DOI
    10.1109/BMEI.2010.5639510
  • Filename
    5639510