Abstract
A hyperthermia simulation system has been developed for the prediction of specific absorption rate(SAR) distribution of object which irradiated by capacitive plate electrodes. A impedance method is used for numerical simulation. Moreover, a temperature distribution of a phantom which irradiated by a 13.56 MHz hyperthermia system is measured by the use of a thermometry system. Results show that there are no significant difference between measured and estimated SAR distribution in this study.
Keywords
biomedical equipment; cancer; dosimetry; electrodes; hyperthermia; numerical analysis; phantoms; temperature measurement; capacitive applicators; capacitive plate electrodes; frequency 13.56 MHz; hyperthermia simulation system; impedance method; phantom temperature distribution; specific absorption rate distribution; thermometry system; Absorption; Applicators; Electrodes; Frequency control; Materials; Specific absorption rate;