• DocumentCode
    544271
  • Title

    A SAR distribution of 13.56 MHz capacitive applicators — Experiment and numerical analysis

  • Author

    Su, Jenn-Lung ; Shih, Yuan-Hsing ; Hsu, Wen-Lin

  • Author_Institution
    Inst. of Biomed. Eng., Chung Yuan Univ., Chungli, Taiwan
  • Volume
    1
  • fYear
    1992
  • fDate
    Oct. 29 1992-Nov. 1 1992
  • Firstpage
    240
  • Lastpage
    241
  • Abstract
    A hyperthermia simulation system has been developed for the prediction of specific absorption rate(SAR) distribution of object which irradiated by capacitive plate electrodes. A impedance method is used for numerical simulation. Moreover, a temperature distribution of a phantom which irradiated by a 13.56 MHz hyperthermia system is measured by the use of a thermometry system. Results show that there are no significant difference between measured and estimated SAR distribution in this study.
  • Keywords
    biomedical equipment; cancer; dosimetry; electrodes; hyperthermia; numerical analysis; phantoms; temperature measurement; capacitive applicators; capacitive plate electrodes; frequency 13.56 MHz; hyperthermia simulation system; impedance method; phantom temperature distribution; specific absorption rate distribution; thermometry system; Absorption; Applicators; Electrodes; Frequency control; Materials; Specific absorption rate;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
  • Conference_Location
    Paris
  • Print_ISBN
    0-7803-0785-2
  • Electronic_ISBN
    0-7803-0816-6
  • Type

    conf

  • DOI
    10.1109/IEMBS.1992.5760945
  • Filename