    Basic considerations of ultrasonic air-bubble detectors for blood extracorporeal circulation

  • Author

    Costa, Luiz Eduardo G da ; Machado, João C.

  • Author_Institution
    Biomedicai Engineering Program - COPPE/Federal University of Rio de Janeiro, P.O. BOX 68510 - 21945 Rio de Janeiro, R.J. Brazil
  • Volume
    2
  • fYear
    1992
  • fDate
    Oct. 29 1992-Nov. 1 1992
  • Firstpage
    380
  • Lastpage
    381
  • Abstract
    This work presents an analysis about the limitation of ultrasonic air-bubble detectors based on the measurement of the wave attenuation caused by the presence of air-bubbles in the blood circuit. Considering as a lower limit the amount of 0.5 ml of air in 1 liter of blood for a blood flow rate of 250 ml/min then it is shown that for ultrasonic frequencies of 50 kHz, 100 kHz, 1 MHz and 5 MHz the detectors may fail if the air-bubble diameter exceeds 110, 74, 44 and 38 μm, respectively.
  • Keywords
    Acoustics; Biomedical measurements; Blood; Frequency measurement; Ultrasonic variables measurement;