• DocumentCode
    544465
  • Title

    Accurate calibration of pneumotacographs using a syringe and polynomial curve fitting

  • Author

    Neto, Antonio Giannella ; Vale, Marcio J O do ; Melo, Marcos F.Vidal

  • Author_Institution
    Biomedicai Engineering Program, COPPE, Federal University of Rio de Janeiro, PO Box 68510, 21945 Rio de Janeiro, RJ, Brazil
  • Volume
    2
  • fYear
    1992
  • fDate
    Oct. 29 1992-Nov. 1 1992
  • Firstpage
    693
  • Lastpage
    694
  • Abstract
    This work presents a simple, reliable and accurate calibration technique for pneumotacographs. It is based on a polynomial curve fitting which minimizes the calculated volume error for a set of injections of a calibrated syringe. In conditions similar to human spirometry the results meet the standards of the American Thoracic Society.
  • Keywords
    Converters; Fitting;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
  • Conference_Location
    Paris, France
  • Print_ISBN
    0-7803-0785-2
  • Electronic_ISBN
    0-7803-0816-6
  • Type

    conf

  • DOI
    10.1109/IEMBS.1992.5761174
  • Filename
    5761174