DocumentCode :
544669
Title :
Microprocessor controlled bioimpedance measurements
Author :
Schmukler, Robert E. ; Dixon, Jonathan W.
Author_Institution :
Electrophysics Branch, Center for Devices and Radiological Health, OST, FDA 12721 Twinbrook Parkway, Rockville, MD 20857
Volume :
6
fYear :
1992
fDate :
Oct. 29 1992-Nov. 1 1992
Firstpage :
2381
Lastpage :
2381
Abstract :
This paper describes the measurements made possible by using microprocessor-control to operate a Solartron 1260 Impedance Analyzer. Calibration experiments were able to demonstrate that an apparent two time-constant impedance measurement is really only a one time-constant system. This is due to artifacts introduced by the parasitic capacitance and the source resistance to the reference plane of the measurement system.