Title :
Video endoscopy: Our experience with image quality assessment
Author :
McMahon, B.P. ; Hegarty, F. ; Malone, J.F.
Author_Institution :
Dept. of Med. Phys. & Bio-Eng., St. James Hosp., Dublin, Ireland
fDate :
Oct. 29 1992-Nov. 1 1992
Abstract :
With over two years experience in the support of a busy endoscopy unit the authors have gathered important information which brings into question the supplier recommendations for set up and calibration of video endoscopic systems . Errors in colour rendition and resolution due to electronic drift, subjective set up, user errors and variation between different imaging systems suggest the need for quality assurance procedures and the development of a set of standards.
Keywords :
calibration; endoscopes; quality assurance; calibration; colour rendition errors; electronic drift; image quality assessment; quality assurance procedures; user errors; video endoscopy; Biology; Switches;
Conference_Titel :
Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
Conference_Location :
Paris
Print_ISBN :
0-7803-0785-2
Electronic_ISBN :
0-7803-0816-6
DOI :
10.1109/IEMBS.1992.5761436