• DocumentCode
    544765
  • Title

    Noninvasive blood pressure measurements on the temporal artery

  • Author

    von Mateahn, Wolf W. ; Buckey, Jay ; Shenoy, Damodar

  • Author_Institution
    Joint Biomedical Engineering Program, University of Texas at Arlington and University of Texas Southwestern Medical Center at Dallas
  • Volume
    6
  • fYear
    1992
  • fDate
    Oct. 29 1992-Nov. 1 1992
  • Firstpage
    2425
  • Lastpage
    2426
  • Abstract
    Systolic and diastolic blood pressures were recorded noninvasively on the temporal artery and compared with systolic and diastolic blood pressures obtained from the brachial artery. Both types of measurements were based on the detection of Korotkoff sounds. Data was collected from five subjects with the subjects lying in a supine position, sitting at rest. and sitting immediately after moderate exercise. This comparative study showed that both pressure measurements track each other within a difference margin of ±10%. which is comparable to the generally accepted accuracy of noninvasive blood pressure measurements.
  • Keywords
    Arteries; Blood; Extraterrestrial measurements; Frequency measurement; Position measurement; Pressure measurement; Skin;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
  • Conference_Location
    Paris, France
  • Print_ISBN
    0-7803-0785-2
  • Electronic_ISBN
    0-7803-0816-6
  • Type

    conf

  • DOI
    10.1109/IEMBS.1992.5761523
  • Filename
    5761523