    Usually, using high and low energy x-ray images we can decompose into soft tissue image and bone image. That is based on that different materials have different attenuation coefficient ratios of high energy x-ray to low energy x-ray. We apply this principle to different materials in soft tissues with different average atomic number, it can further separate image of different tissues such as lung tissue and connective tissue like bronchus. This method has been proved by experiment and it has potential use for identifying the composition of soft tissues.