• DocumentCode
    546187
  • Title

    Accuracy investigation of an ultra-wideband time domain microwave imaging system

  • Author

    Zeng, X. ; Fhager, A. ; Linner, P. ; Persson, M. ; Zirath, H.

  • Author_Institution
    Dept. of Signals & Syst., Chalmers Univ. of Technol., Gothenburg, Sweden
  • fYear
    2011
  • fDate
    11-15 April 2011
  • Firstpage
    1928
  • Lastpage
    1932
  • Abstract
    We investigate the measurement accuracy of an ultra-wideband time domain microwave tomography system. A phantom is imaged by means of the system and the reconstructed images are compared to those obtained with an ultra-wideband frequency domain system. The results suggest that the permittivity reconstructions obtained by means of the time domain and the frequency domain systems are very similar to each other, but the conductivity profile reconstructed from time domain measurements presents worse quality than that obtained from frequency domain measurements.
  • Keywords
    image reconstruction; microwave imaging; phantoms; time-frequency analysis; ultra wideband communication; frequency domain system; image reconstruction; measurement accuracy; microwave imaging; microwave tomography system; phantom; time-domain analysis; ultra wideband; Frequency domain analysis; Image reconstruction; Microwave imaging; Microwave measurements; Microwave theory and techniques; Time domain analysis; Time measurement;