• DocumentCode
    55851
  • Title

    A Low-Operating-Voltage Wireless Intermediate-Range Scheme for Energy and Signal Transmission by Magnetic Coupling for Implantable Devices

  • Author

    Sai Chun Tang

  • Author_Institution
    Med. Sch., Radiol. Dept., Harvard Univ., Boston, MA, USA
  • Volume
    3
  • Issue
    1
  • fYear
    2015
  • fDate
    Mar-15
  • Firstpage
    242
  • Lastpage
    251
  • Abstract
    The increasing sophistication and power demands of medical implantable devices have motivated a variety of research on wireless energy transmission, aiming to provide higher power capability, wider receiver position, and improved performance, including safety and efficiency. This paper demonstrates a low-operating-voltage wireless energy and signal transmission method to achieve these aims. We developed a 30-cm transmitting coil in a Helmholtz-coil configuration. Innovative coil segmentation and distributive resonance techniques are employed to substantially reduce the excitation voltage and the voltage over the coil. Magnetic field simulations show that the magnetic field generated by the scheme is uniform and applicable in a wide region bounded by the coils. The receiving coil has three turns and a diameter of 2 cm. The receiver can deliver 350 mW when the transmitting coil excitation is 1 Arms/6.68 Vrms and the maximum voltage over the coil is 32 Vrms, which is much lower than that required in the most up-to-date method (5 A/3.5 kV). This method can provide sufficient power to operate many kinds of medical implants, especially deep-seated and locomotive devices, such as capsule endoscopes. An operating frequency of 6.1 MHz is chosen. At such a high frequency, the receiving coil does not need ferromagnetic core so magnetic resonance imaging compatible implants can be accomplished.
  • Keywords
    biomedical MRI; endoscopes; ferromagnetic materials; inductive power transmission; magnetic fields; prosthetics; Helmholtz-coil configuration; capsule endoscopes; coil segmentation; current 5 A; deep-seated devices; distributive resonance techniques; ferromagnetic core; frequency 6.1 MHz; locomotive devices; magnetic coupling; magnetic field simulations; magnetic resonance imaging; medical implantable devices; medical implants; power 350 mW; signal transmission; size 2 cm; size 30 cm; voltage 3.5 kV; wireless energy transmission; Coils; Eddy currents; Endoscopes; Implants; Magnetic fields; Magnetic resonance imaging; Wireless communication; Coil segmentation; implantable medical devices; low-voltage operation; wireless intermediate-range scheme for energy and signal transmission (WISEST);
  • fLanguage
    English
  • Journal_Title
    Emerging and Selected Topics in Power Electronics, IEEE Journal of
  • Publisher
    ieee
  • ISSN
    2168-6777
  • Type

    jour

  • DOI
    10.1109/JESTPE.2014.2314388
  • Filename
    6780635