Title :
Slice scanning mode for traveling wave MPI
Author :
Vogel, P. ; Ruckert, M.A. ; Kullmann, W.H. ; Jakob, P.M. ; Behr, V.C.
Author_Institution :
Dept. of Exp. Phys. (Biophys.), Univ. of Wurzburg, Wurzburg, Germany
Abstract :
The sample shown in fig. 1B contains a tube with an inner diameter of 860 μm filled with undiluted Resovist® (Bayer-Schering, Germany) embedded in acrylic glass. The acquisition time for one slice is tslice=1/f1, with 4000 averages and f1=920 Hz the overall acquisition time is t=4000/f1=4,35 s. The intrinsic spatial resolution is given by the setup and was calculated to be 2 mm in z- and 4 mm in x-direction. The deconvoluted image (see fig. 1B) shows a good agreement and proves the feasibility of imaging studies, e.g. in pre-clinical context, based on TWMPI.
Keywords :
biomagnetism; biomedical MRI; deconvolution; glass; magnetic particles; acquisition time; acrylic glass; deconvoluted image; intrinsic spatial resolution; preclinical context; size 860 mum; slice scanning mode; traveling wave MPI; undiluted Resovist; x-direction; Coils; Educational institutions; Imaging; Magnetic particles; Physics; Receivers; Trajectory;
Conference_Titel :
Magnetic Particle Imaging (IWMPI), 2013 International Workshop on
Conference_Location :
Berkeley, CA
Print_ISBN :
978-1-4673-5520-9
Electronic_ISBN :
978-1-4673-5521-6
DOI :
10.1109/IWMPI.2013.6528369