• DocumentCode
    609673
  • Title

    An online recursive ICA based real-time multichannel EEG system on chip design with automatic eye blink artifact rejection

  • Author

    Jui-Chieh Liao ; Wei-Yeh Shih ; Kuan-Ju Huang ; Wai-Chi Fang

  • Author_Institution
    Dept. of Electron. Eng., Nat. Chiao Tung Univ., Hsinchu, Taiwan
  • fYear
    2013
  • fDate
    22-24 April 2013
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    This paper presents an online recursive ICA (ORICA) based real-time multi-channel EEG system on chip design with automatic eye blink artifact rejection. Since EEG signals are very feeble, they are easy to be contaminated by artifacts. Among all artifacts, eye blink artifact dose the most significant harm to EEG signals. For acquiring artifact free EEG signals, ICA is a popular method for artifacts extraction. After extraction of ICA, eye blink artifacts are rejected to improve the reliability of EEG applications such as brain computer interfaces (BCIs). To promote the feasibility and convenience of BCIs, a real-time ICA algorithm, ORICA, is adopted in this system. With ORICA, the system immediately finishes each ICA result after each sample time. After each ICA result is generated, the automatic eye blink artifact rejection and inverse ICA are performed to acquire eye blink artifact free EEG signals in realtime. For system portability and high integration, a front-end interface of commercial IC, ADS1298 provided by TI, and Bluetooth interface, UART, are implemented. The system is designed used TSMC 90nm CMOS technology with 8 channels EEG processing in 128 Hz sample rate of raw data and consumes 8.56 mW at 50 MHz clock rate. The design methods of the proposed EEG system are provided in this paper. The performance and processing results of the system are also shown to reach 0.2532 s latency after each EEG sample time.
  • Keywords
    CMOS digital integrated circuits; electroencephalography; integrated circuit design; medical signal processing; real-time systems; recursive estimation; signal reconstruction; ADS1298; Bluetooth interface; EEG sample time; TI; TSMC CMOS technology; UART; automatic eye blink artifact rejection; chip design; clock rate; commercial IC; department real-time multichannel EEG system-based ORICA; frequency 128 Hz; frequency 50 MHz; front-end interface; online recursive ICA based real-time multichannel EEG system; power 8.56 mW; system portability; Algorithm design and analysis; Covariance matrices; Design methodology; Electroencephalography; Real-time systems; System-on-chip; Training;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    VLSI Design, Automation, and Test (VLSI-DAT), 2013 International Symposium on
  • Conference_Location
    Hsinchu
  • Print_ISBN
    978-1-4673-4435-7
  • Type

    conf

  • DOI
    10.1109/VLDI-DAT.2013.6533850
  • Filename
    6533850