DocumentCode :
636556
Title :
An implantable wireless system for muscle afferent recording from the sciatic nerve during functional electrical stimulation
Author :
Kang-Il Song ; Shon, Ahnsei ; Jun-Uk Chu ; Kuiwon Choi ; Dosik Hwang ; Inchan Youn
Author_Institution :
Biomed. Res. Inst., Korea Inst. of Sci. & Technol., Seoul, South Korea
fYear :
2013
fDate :
3-7 July 2013
Firstpage :
3610
Lastpage :
3613
Abstract :
An implantable wireless system was developed for recording muscle afferent activity and stimulating peripheral nerves with cuff electrodes. The proposed system was fabricated into the nerve cuff electrode, neural amplifier, neural stimulator, and wireless communication system with battery power. The nerve cuff electrode and neural amplifier were designed to improve the signal-to-interference ratio and signal-to-noise ratio. The wireless communication system was designed based on the medical implant communication service regulations to be suitable for implantation. The main function of this system was to extract muscle afferent activity from peripheral nerve during functional electrical stimulation. The cuff electrodes were chronically implanted on the sciatic nerve for recording and on the tibial and peroneal nerves for stimulation. When the extension and flexion movements of ankle joint were elicited from alternative electrical stimuli, the corresponding neural signals and ankle angles were recorded simultaneously. The muscle afferent activity was then extracted from the recorded neural signal through a simple blanking process. The experimental results showed that the ankle movements could be detected from the extracted muscle afferent activity.
Keywords :
biomechanics; biomedical electrodes; feature extraction; medical signal detection; neural nets; neuromuscular stimulation; patient diagnosis; prosthetics; wireless LAN; alternative electrical stimuli; ankle joint extension movement; ankle joint flexion movement; battery power; feature extraction; functional electrical stimulation; implantable wireless system; medical implant communication service regulations; muscle afferent activity recording; nerve cuff electrode; neural amplifier; neural signal; neural stimulator; peripheral nerve stimulation; peroneal nerves; sciatic nerve; signal-to-interference ratio; signal-to-noise ratio; simple blanking process; tibial nerves; wireless communication system; Electrodes; Implants; Muscles; Neuromuscular stimulation; Signal to noise ratio; Wireless communication;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2013 35th Annual International Conference of the IEEE
Conference_Location :
Osaka
ISSN :
1557-170X
Type :
conf
DOI :
10.1109/EMBC.2013.6610324
Filename :
6610324
Link To Document :
بازگشت