DocumentCode
644743
Title
Calibration of mm-wave power meters using a broadband calorimeter in the frequency range from 110 GHz to 170 GHz
Author
Shimaoka, Kazuhiro ; Kinoshita, Moto ; Fujii, Kenichi ; Tosaka, T.
Author_Institution
Nat. Inst. of Adv. Ind. Sci. & Technol., Tsukuba, Japan
fYear
2013
fDate
1-6 Sept. 2013
Firstpage
1
Lastpage
2
Abstract
A broadband rectangular waveguide calorimeter is fabricated to establish a reliable mm-wave power standard in the frequency range from 110 GHz to 170 GHz. Two types of mm-wave power meter are calibrated using this calorimeter and their calibration factors are reported with their uncertainties. Uncertainty calculations are based on the Monte Carlo method.
Keywords
Monte Carlo methods; calibration; calorimeters; measurement uncertainty; microwave measurement; power meters; rectangular waveguides; Monte Carlo method; broadband calorimeter; broadband rectangular waveguide calorimeter; frequency 110 GHz to 170 GHz; millimeter wave power meter calibration; millimeter wave power standard; uncertainty calculation; Broadband communication; Calibration; Frequency measurement; Power measurement; Time measurement; Transmission line measurements; Uncertainty;
fLanguage
English
Publisher
ieee
Conference_Titel
Infrared, Millimeter, and Terahertz Waves (IRMMW-THz), 2013 38th International Conference on
Conference_Location
Mainz
Type
conf
DOI
10.1109/IRMMW-THz.2013.6665712
Filename
6665712
Link To Document