• DocumentCode
    657229
  • Title

    Cu2O based electrochemical sensor for direct glucose detection

  • Author

    Majzlikova, Petra ; Prasek, Jan ; Chomoucka, Jana ; Drbohlavova, Jana ; Pekarek, Jan ; Hrdy, Radim ; Hubalek, J. ; Trnkova, Libuse

  • Author_Institution
    Dept. of Microelectron., Brno Univ. of Technol. Brno, Brno, Czech Republic
  • fYear
    2013
  • fDate
    3-6 Nov. 2013
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    Two types of working electrodes of electrochemical sensors were fabricated using screen-printing and spray-coating deposition methods and previously synthetized Cu2O micro/nanoparticles and used for direct electrochemical glucose detection. Glucose was successfully electrochemically detected against conventional Ag/AgCl reference electrode and platinum auxiliary electrode at potential of 0.5 V and 0.75 V for spray-coated and screen-printed electrode, respectively. Better sensitivity was observed using spray-coated electrode. For both electrodes types we obtained linear calibration curve in physiological range of concentrations from 0.5 to 10 mmol/L. Furthermore, the limit of detection lower than 100 μmol/L was achieved.
  • Keywords
    II-VI semiconductors; biochemistry; biomedical electrodes; blood; calibration; copper compounds; electrochemical electrodes; electrochemical sensors; nanomedicine; nanoparticles; spray coating techniques; wide band gap semiconductors; Cu2O; direct glucose detection; electrochemical sensor; electrodes; linear calibration curve; microparticles; nanoparticles; screen-printing deposition; spray-coating deposition; voltage 0.5 V; voltage 0.75 V; Biosensors; Calibration; Electric potential; Electrodes; Nanoparticles; Sensitivity; Sugar;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    SENSORS, 2013 IEEE
  • Conference_Location
    Baltimore, MD
  • ISSN
    1930-0395
  • Type

    conf

  • DOI
    10.1109/ICSENS.2013.6688515
  • Filename
    6688515