DocumentCode
682179
Title
Absolute measurement of low-voltage capacitor voltage coefficient by double voltage method
Author
Li Wenfeng ; Li Dong ; Shao Haiming ; Li Yanqiang ; Zhang Huanghui
Author_Institution
Dept. of Instrum. Sci. & Opto-Electron. Eng., Beijing Inf. Sci. & Technol. Univ., Beijing, China
Volume
1
fYear
2013
fDate
16-19 Aug. 2013
Firstpage
411
Lastpage
415
Abstract
The voltage coefficient of low voltage capacitor is the most important fact that affects the accuracy degree of capacitance value of capacitor. Its value and feature must be measured out when it used in the high accuracy situation. This paper presents an absolute double voltage measure method that gets the absolute voltage coefficient curve from 0 V to 1000 V needs only 5-step measurement. And proposed the related compensate measures and methods against the possible problems in measure system. After compensation, the measure error is only related to the accuracy of current comparator bridge. In this paper the measurement uncertainty is about 1×10-6 when ignore the linear fitting error.
Keywords
capacitors; compensation; voltage measurement; absolute double voltage measurement method; compensation; current comparator bridge; low-voltage capacitor voltage coefficient; measurement uncertainty; Accuracy; Bridge circuits; Capacitance measurement; Capacitors; Current measurement; Measurement uncertainty; Voltage measurement; double voltage method; low-voltage precision capacitor; voltage coefficient; voltage ratio;
fLanguage
English
Publisher
ieee
Conference_Titel
Electronic Measurement & Instruments (ICEMI), 2013 IEEE 11th International Conference on
Conference_Location
Harbin
Print_ISBN
978-1-4799-0757-1
Type
conf
DOI
10.1109/ICEMI.2013.6743009