• DocumentCode
    68291
  • Title

    Marker-Based Surgical Instrument Tracking Using Dual Kinect Sensors

  • Author

    Hongliang Ren ; Wei Liu ; Lim, Andrew

  • Author_Institution
    Dept. of Biomed. Eng., Nat. Univ. of Singapore, Singapore, Singapore
  • Volume
    11
  • Issue
    3
  • fYear
    2014
  • fDate
    Jul-14
  • Firstpage
    921
  • Lastpage
    924
  • Abstract
    This paper presents a passive-marker-based optical tracking system utilizing dual Kinect sensors and additional custom optical tracking components. To obtain sub-millimeter tracking accuracy, we introduce robust calibration of dual infrared sensors and point correspondence establishment in a stereo configuration. The 3D localization is subsequently accomplished using multiple back projection lines. The proposed system extends existing inexpensive consumer electronic devices, implements tracking algorithms, and shows the feasibility of applying the proposed low-cost system to surgical training for computer assisted surgeries.
  • Keywords
    biomedical equipment; calibration; infrared detectors; optical tracking; surgery; calibration; computer assisted surgeries; dual infrared sensors; dual kinect sensors; marker-based surgical instrument tracking; multiple back projection lines; passive-marker-based optical tracking system; submillimeter tracking accuracy; Accuracy; Calibration; Cameras; Instruments; Sensors; Surgery; Three-dimensional displays; Kinect sensor; navigated surgery; optical tracking; tracking;