• DocumentCode
    69826
  • Title

    Drop Measurement System for Biomedical Application

  • Author

    Norgia, Michele ; Magnani, Alessandro ; Melchionni, Dario ; Pesatori, Alessandro

  • Author_Institution
    Dipt. di Elettron., Inf. e Bioing., Politec. di Milano, Milan, Italy
  • Volume
    64
  • Issue
    9
  • fYear
    2015
  • fDate
    Sept. 2015
  • Firstpage
    2513
  • Lastpage
    2517
  • Abstract
    This paper describes the realization of a sensor capable of measuring the volume of a drop in free fall. The sensor realized is made of a simple low-cost red laser and a photodiode and optics to focus the beam on the light sensor. In this way, when the drop is falling down, it is possible to estimate its volume from the signal generated on the photodiode while getting through the laser beam. This system may be very useful together with a drip chamber to have accurate and low-cost volume and flow rate measurements of the infused substances.
  • Keywords
    drops; laser applications in medicine; optical sensors; volume measurement; biomedical application; drip chamber; drop volume measurement; flow rate measurements; free fall; light sensor beam focusing; optics; photodiode; red laser; Biomedical measurement; Biomedical optical imaging; Laser beams; Measurement by laser beam; Optical sensors; Photodiodes; Volume measurement; Biomedical monitoring; fluid flow measurement; laser measurement application; optoelectronic devices; volume measurement; volume measurement.;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2015.2398972
  • Filename
    7042930