    Flow cytometer for reference measurements of blood cell concentrations with low uncertainty

  • Author

    Kammel, M. ; Kummrow, A. ; John, M. ; Reitz, S. ; Witt, K. ; Neukammer, J.

  • Author_Institution
    Phys.-Tech. Bundesanstalt, Berlin, Germany
  • fYear
    2015
  • fDate
    7-9 May 2015
  • Firstpage
    517
  • Lastpage
    520
  • Abstract
    A method is described to determine cell concentrations for erythrocytes and leukocytes with typical uncertainties of 0.8% and 2%, respectively. The unique method is designed in such a way that results are traceable to SI derived units. The sample suspensions are prepared under gravimetric control. The sample volume measured in cell counting is determined by weighing, including separate density measurements. Random coincidences are a significant challenge in particle counting leading to underestimated cell concentrations. In the method described here this challenge is met by recording the integrated dead time in particle counting and applying dilution series. This method is now routinely applied to provide reference measurement values for external quality assurance of German hematology laboratories for the complete blood count.
  • Keywords
    biomedical measurement; blood; cellular biophysics; density measurement; weighing; German hematology laboratory; SI derived unit; blood cell concentration reference measurement; blood count; cell concentration determination; cell counting; density measurement; dilution series; erythrocyte; flow cytometer; gravimetric control; leukocyte; low uncertainty; particle counting; reference measurement value; Atmospheric measurements; Blood; Impedance; Loss measurement; Particle measurements; Uncertainty; Volume measurement; cell concentration; dead time correction; erythrocytes; external quality assurance; leukocytes; reference measurement procedure;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Medical Measurements and Applications (MeMeA), 2015 IEEE International Symposium on
  • Conference_Location
    Turin
  • Type

    conf

  • DOI
    10.1109/MeMeA.2015.7145258
  • Filename
    7145258