• DocumentCode
    757602
  • Title

    Scintillator and CMOS APS Imager for Radiography Conditions

  • Author

    Kim, Kwang Hyun ; Kim, Young Soo

  • Author_Institution
    Coll. of Dentistry, Chosun Univ., Gwangju
  • Volume
    55
  • Issue
    3
  • fYear
    2008
  • fDate
    6/1/2008 12:00:00 AM
  • Firstpage
    1327
  • Lastpage
    1332
  • Abstract
    We evaluated X-ray image performance for several scintillators and a CMOS APS imager by both diagnostic radiography and mammography conditions. Commercially available scintillators such as Lanex screen, needle structured CsI (Tl), and fiber optic structured CsI (Tl) were coupled with a CMOS APS imager. The X-ray machines used in this study were fixed tube voltage of 80 kVp and variable tube current over 0.33 mA using LISTEMtrade for diagnostic radiography, and fixed tube voltage of 28 kVp and variable tube current over 16 mA using Alpha-STtrade for mammography. We used the RadEye1trade CMOS APS imager having an active area of 25 mm by 50 mm with the pixel size of 48 m.
  • Keywords
    CMOS image sensors; diagnostic radiography; mammography; solid scintillation detectors; Alpha-STtrade; CMOS APS imager; LISTEMtrade; Lanex screen scintillator; RadEye1trade CMOS APS imager; diagnostic radiography; fiber optic structured CsI (Tl) scintillator; mammography; needle structured CsI (Tl) scintillator; Amorphous materials; CMOS image sensors; Diagnostic radiography; Mammography; Needles; Optical fibers; Pixel; X-ray detection; X-ray detectors; X-ray imaging; CMOS active pixel sensor; diagnostic radiography; mammography; scintillators;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2008.922214
  • Filename
    4545142