DocumentCode
75941
Title
Comparison of Cephalic and Extracephalic Montages for Transcranial Direct Current Stimulation—A Numerical Study
Author
Noetscher, G.M. ; Yanamadala, J. ; Makarov, Sergey N. ; Pascual-Leone, A.
Author_Institution
Dept. of Electr. & Comput. Eng., Worcester Polytech. Inst., Worcester, MA, USA
Volume
61
Issue
9
fYear
2014
fDate
Sept. 2014
Firstpage
2488
Lastpage
2498
Abstract
While studies have shown that the application of transcranial direct current stimulation (tDCS) has been beneficial in the stimulation of cortical activity and treatment of neurological disorders in humans, open questions remain regarding the placement of electrodes for optimal targeting of currents for a given functional area. Given the difficulty of obtaining in vivo measurements of current density, modeling of conventional and alternative electrode montages via the finite element method has been utilized to provide insight into tDCS montage performance. It has been shown that extracephalic montages might create larger total current densities in deeper brain regions, specifically in white matter as compared to an equivalent cephalic montage. Extracephalic montages might also create larger average vertical current densities in the primary motor cortex and in the somatosensory cortex. At the same time, the horizontal current density either remains approximately the same or decreases. The metrics used in this paper include either the total local current density through the entire brain volume or the average vertical current density as well as the average horizontal current density for every individual lobe/cortex.
Keywords
bioelectric potentials; biomedical electrodes; biomedical measurement; brain models; current density; finite element analysis; neurophysiology; patient treatment; alternative electrode montage modeling; average horizontal current density; average vertical current density; brain volume; conventional electrode montage modeling; cortical activity stimulation; current density metrics; deep brain regions; electrode cephalic montages; electrode extracephalic montages; electrode placement; finite element method; functional brain area; in vivo current density measurements; lobe; neurological disorder treatment; numerical study; optimal current targeting; primary motor cortex; somatosensory cortex; tDCS application; tDCS montage performance; total local current density; transcranial direct current stimulation; white matter; Brain modeling; Current density; Electrodes; Finite element analysis; Image segmentation; Numerical models; Skin; In vivo current density; transcranial direct current stimulation (tDCS);
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/TBME.2014.2322774
Filename
6847106
Link To Document