Title :
Automatic Hysteresis Measurement System with High Accuracy
Author :
Ueda, F. ; Takeda, K. ; Yamaguchi, T.
Author_Institution :
Fukuoka Institute of Technology, Fukuoka.
Abstract :
A newly developed microcomputer-controlled device for automatic high-precision hysteresis measurement is reported. Microcomputer integration of dB/dt and dH/dt signals can proceed by 1) direct accumulation of the differential voltages, 2) analog integration through use of a mirror integrator, or 3) digital integration using a V - f converter. Barkhausen noise in signals leads to problems in adopting 1), and other problems attend 2), so the authors employed 3) for integration of both B and H signals. This enabled highly precise measurement of the static magnetization characteristics of single sheet samples, with a stability of 3 à 10¿6 Wb·turn/ min or more.
Keywords :
Analog-digital conversion; Integral equations; Magnetic hysteresis; Magnetization; Microcomputers; Mirrors; Permeability measurement; Stability; Tellurium; Voltage;
Journal_Title :
Magnetics in Japan, IEEE Translation Journal on
DOI :
10.1109/TJMJ.1985.4549001