    Phase Detection Electromagnetic Flowmeter-Design and Use

  • Author

    Olmsted, Frederick

  • Author_Institution
    Research Division, Cleveland Clinic Foundation, and Frank E. Bunts Educational Institute, Cleveland, Ohio
  • Volume
    9
  • Issue
    2
  • fYear
    1962
  • fDate
    4/1/1962 12:00:00 AM
  • Firstpage
    88
  • Lastpage
    92
  • Abstract
    Two types of gated sine-wave electromagnetic flowmeter are examined critically, and found to be sensitive to phase shift in the transducer signal. This phase shift is always found in a sine-wave flow transducer because of imperfection in construction, particularly in those used for physiology and medicine. A flowmeter design that discards amplitude modulation and is sensitive to phase change only is described. Advantages over other types of electromagnetic flowmeter are discussed. Results obtained with the phase detection flowmeter are illustrated.
  • Keywords
    Blood; Capacitance; Electrodes; Electromagnetic shielding; Force measurement; Magnetic fields; Magnetic separation; Magnetic shielding; Phase detection; Transducers;