• DocumentCode
    783299
  • Title

    Large, Multiple Element Germanium Spectrometers

  • Author

    Lalovic, B. ; Azuma, R.E. ; Petrovic, B.

  • Author_Institution
    University of Toronto Toronto, Ontario, Canada
  • Volume
    14
  • Issue
    1
  • fYear
    1967
  • Firstpage
    514
  • Lastpage
    520
  • Abstract
    Large coaxial Ge(Li) diodes have been assembled in multiple arrays in one cryostat in order to provide spectrometers of very large sensitive volumes. A double element and a four element spectrometer are described in detail. The latter spectrometer has a total sensitive volume of 180 cm3. The efficiencies and resolutions of the spectrometers were neasured for various combinations of diodes and summing procedures . In the optimium arrangement for the four element spectrometer, a resolution of 15 keV at 1.33 MeV was obtained, and the full energy efficiency was found to be equal to one quarter of that for a 3" × 3" NaI (Tl) scintillation counter. An efficiency curve for the two element spectrometer for energies up to 10 Mev is presented.
  • Keywords
    Assembly; Coaxial components; Diodes; Energy resolution; Germanium; Protons; Reactive power; Solid scintillation detectors; Spectroscopy; Stationary state;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1967.4324463
  • Filename