DocumentCode :
785305
Title :
Dosimeter Properties of Electret Ionization Chambers
Author :
Pretzsch, G. ; Darschel, B. ; Schanmuth, T.
Author_Institution :
Technische Universitÿt Dresden Sektion Physik Dresden German Democratic Republic
Issue :
3
fYear :
1986
fDate :
6/1/1986 12:00:00 AM
Firstpage :
437
Lastpage :
441
Abstract :
An integrating electret dosimeter is an analog of a common ionization chamber, replacing one electrode by an electret. Using such dosimeters for personnel and environmental monitoring the radiation registration sensitivity as a function of kind and energy of radiation, as well as of wall material and interelectrode distance, is of interest. In the present paper these values were determined theoretically and experimentally for incident X-and gamma rays and neutrons as well as for radioactive gases (tritium, radon) for a parallel-plate chamber. Furthermore, the radiation registration properties of an electret pulse chamber were studied. The chamber contains a coronacharged PTFE electret. In order to get high charge stability, electret formation at elevated temperatures was studied.
Keywords :
Electrets; Electrodes; Gamma rays; Gases; Ionization chambers; Neutrons; Personnel; Radiation monitoring; Radioactive materials; Stability;
fLanguage :
English
Journal_Title :
Electrical Insulation, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9367
Type :
jour
DOI :
10.1109/TEI.1986.349089
Filename :
4157002
Link To Document :
بازگشت