DocumentCode :
787735
Title :
The Channel Electron Multiplier as an Alpha Detector
Author :
Kaiser, W.C. ; Managan, W.W. ; Mayer, T.A.
Author_Institution :
Electronics Division Argonne National Laboratory Argonne, Illinois 60439
Volume :
15
Issue :
1
fYear :
1968
Firstpage :
498
Lastpage :
502
Abstract :
The response of Bendix CEM 401 channel electron multipliers to a hot filament electron source and a 244Cm alpha source has been investigated. This investigation shows that a CEM operated in the saturated mode produces output pulses for alpha particles which are, on the average, about twice the height of those for single electrons. A series of pulse height distribution curves obtained with various alpha absorbers and electric field potentials between the source and CEM are presented to substantiate this conclusion.
Keywords :
Alpha particles; Assembly; Attenuators; Costs; Detectors; Diffraction; Electron multipliers; Helium; Neutrons; Powders;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1968.4324892
Filename :
4324892
Link To Document :