DocumentCode :
790367
Title :
Variations in Single Photon Pulse Height Distribution with Focusing Voltage in Photomultipliers
Author :
Inman, F.W. ; Muray, J.J.
Author_Institution :
Dept. of Physics, Mankato State College Mankato, Minnesota
Volume :
16
Issue :
2
fYear :
1969
fDate :
4/1/1969 12:00:00 AM
Firstpage :
62
Lastpage :
64
Abstract :
In photomultipliers the focusing voltage which gives the most peaked distribution is also the one which results in the largest count rate. For low focusing voltage there is a significant loss in single-photon events. The results of this experiment are consistent with the idea that the pulse distribution depends upon variations in the effective secondary emission ratio.
Keywords :
Cathodes; Electrodes; Electron emission; Exponential distribution; Low voltage; Manufacturing; Optimized production technology; Photomultipliers; Physics; Shape;
fLanguage :
English