Detection of Low Energy X Rays with Si(Li) Detectors
Author
Jaklevic, J.M. ; Goulding, F.S.
Author_Institution
Lawrence Radiation Laboratory University of California Berkeley, California
Volume
18
Issue
1
fYear
1971
Firstpage
187
Lastpage
191
Abstract
The continuing improvement in energy resolution of semiconductor detector X-ray spectrometers has led to interest in the use of these devices at energies less than 2 keV. This is an energy region of potential analytical interest since the K X rays of several elements of biological and chemical importance occur at these energies. However, the low X-ray fluorescence yield in low-Z elements, combined with the absorption of low-energy X rays due to the entry window of the vacuum chanber and that of the detector, have made work in this energy region impossible with the standard semiconductor detector X-ray spectrometers. In the present work, the X-ray absorption path has been reduced to a minimum to permit low-energy X ray studies of excitation, entry window thickness, detector linearity and resolution. Using electron beam excitation on low-Z targets, we have performed measurements of characteristic K X rays of elements down to and including carbon (277 eV).
Keywords
Biochemical analysis; Chemical analysis; Chemical elements; Electromagnetic wave absorption; Energy resolution; Fluorescence; Potential energy; Spectroscopy; X-ray detection; X-ray detectors;