Title :
Linac X-Ray and Electron Dosimetry Using a Remote Control Pin Diode
Author_Institution :
Air Force Cambridge Research Laboratories (AFSC) Bedford, Mass. 01730
fDate :
6/1/1971 12:00:00 AM
Abstract :
A remotely positioned PIN diode was built and has been used for x-ray and electron dosimetry. The data for a plot of dose vs distance (1 meter) for various degrading foils can be obtained for 40 or 50 data points in a few minutes. These data can, in turn, be used for rapid predetermined changes in the Linac dose rate without retuning the accelerator.
Keywords :
Aluminum; Calibration; Circuits; Degradation; Diodes; Dosimetry; Electron beams; Linear particle accelerator; Probes; Tungsten;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :