DocumentCode
804113
Title
The Direct Evaluation of Low Level Counting Techniques Employing Ge(Li) Detectors
Author
Walford, G.V. ; Aliaga-Kelly, D.T.W. ; Gilboy, W.B.
Author_Institution
Nuclear Enterprises Ltd., Sighthill Industrial Estate, Edinburgh 11, U.K.
Volume
19
Issue
1
fYear
1972
Firstpage
127
Lastpage
134
Abstract
The low level counting capabilities and critical parameters of Ge(Li) gamma ray spectrometers are investigated using an expression for the spectrometer sensitivity limit. The significance of the counting parameters involved and the practical means by which they may be optimised have also been considered. A comparison method is used to evaluate the effectiveness and behaviour of several available counting techniques (e.g. coincidence and anti-coincidence methods).
Keywords
Carbon capture and storage; Detectors; Energy resolution; Equations; Fabrication; Face detection; Gamma rays; Scattering; Shape; Spectroscopy;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1972.4326497
Filename
4326497
Link To Document