DocumentCode :
804611
Title :
Improved Amplitude and Rise Time Compensated Timing with Ge Detectors
Author :
Cho, Z.H. ; Chase, R.L.
Author_Institution :
Brookhaven National Laboratory Upton, N. Y. 11973
Volume :
19
Issue :
1
fYear :
1972
Firstpage :
451
Lastpage :
460
Abstract :
The energy range over which signal timing can be made nearly independent of amplitude and rise time has been extended downward by about an order of magnitude by employing dual priming of a zero crossing discriminator. By introducing another coarsely compensated timer between the high resolution timer and its priming leading edge discriminator, one can optimize the timing accuracy without sacrifice of dynamic range. Time walk of less than 1 ns has been achieved from about 20 keV to over 1 MeV.
Keywords :
Dynamic range; Gamma ray detection; Gamma ray detectors; Laboratories; Pulse shaping methods; Shape; Signal resolution; Solid scintillation detectors; Timing; Voltage;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1972.4326545
Filename :
4326545
Link To Document :
بازگشت